Technical
Specifications of SM-4 |
| Principle
of measurement |
Thermo
Catalytic converter followed by atomic absorption detector
(@ 253.7 nm) |
| Measuring
range |
0-0.1
ug/m3; 0-10 ug/m3; 0-50 ug/m3 |
| Detection
Limit |
0.1
ng/m3 |
| Response
Time |
10
sec … 90 sec |
| Max.Sample
gas temperature |
(600 o F) |
| Ambient
temperature range allowed |
1
o F to 120 o F |
| Sample
gas flow |
40
–60 l/h |
| Sample
line length |
1000 Feet typical |
| Electrical
Power consumption |
110
V AC/50-60 Hz; ca.750 VA plus sample line ca. 130 VA per
meter length |
| Air
consumption |
2
bar (29 PSI); ca. 10 l/hour typical |
| Dimensions
for installation |
Ca.55
x 165 x 80 cm (w x h x d) |
| Weight |
Ca.
50 kg (including mounting frame) |
| Service
interval |
6
months |